Specification of health data transfer from devices to DiGA (§ 374a SGB V)
| Official URL: https://gematik.de/fhir/hddt/StructureDefinition/hddt-blood-glucose-measurement | Version: 0.1.0 | |||
| Draft as of 2025-10-27 | Computable Name: HddtBloodGlucoseMeasurement | |||
Copyright/Legal: Copyright (c) 2025 gematik GmbH |
||||
Profile for capturing blood glucose measurements as FHIR Observation resources.
This profile defines the exchange of a single measurement data for the Mandatory Interoperable Value (MIV) "Blood Glucose Measurement" which is technically defined by the ValueSet hddt-miv-blood-glucose-measurement. This MIV is e.g. implemented by blood glucose meter (glucometer) that can connect to a Personal Health Gateway (e.g. a mobile app for keeping diabetes diary) through wireless or wired communication.
Obligations and Conventions:
Each Blood Glucose Measurement MUST either hold a reference to a Sensor Type And Calibration Status DeviceMetric resource or to a Personal Health Device Device resource (eXclusive OR). A reference to Sensor Type And Calibration Status MUST be provided from the Observation resource if the sensor for measuring blood glucose needs to be calibrated (either automatically or by the user) or if the sensor may change its calibration status over time.
Constraints applied:
status is restricted to finalcode is constrained to the ValueSet that represents the MIV Blood Glucose Measurementeffective[x] is restricted to effectiveDateTime and constrained as mandatory.value[x] is restricted to valueQuantity. The elements valueQuantity.value, valueQuantity.system, and valueQuantity.code are constrained in a way that a value MUST be provided and that UCUM MUST be used for encoding the unit of measurement. Observation.valueQuantity MAY only be omitted in case of an error that accured with the measurement. In this case, Observation.dataAbsentReason MUST be provided.device is set to be mandatory in order to provide the DiGA with information about the sensor’s calibration status and with information about the static and dynamic attributes of the Personal Health Device.Usages:
You can also check for usages in the FHIR IG Statistics
Description of Profiles, Differentials, Snapshots and how the different presentations work.
| Name | Flags | Card. | Type | Description & Constraints Filter: ![]() ![]() |
|---|---|---|---|---|
![]() |
C | 0..* | Observation | Measurements and simple assertions Constraints: obs-6, obs-7 |
![]() ![]() |
?!Σ | 0..1 | uri | A set of rules under which this content was created |
![]() ![]() |
?! | 0..* | Extension | Extensions that cannot be ignored |
![]() ![]() |
?!Σ | 1..1 | code | Measurement status Binding: ObservationStatus (required): Codes providing the status of an observation. Fixed Value: final |
![]() ![]() |
Σ | 1..1 | CodeableConcept | Measurement type of blood glucose Binding: ValueSet - Blood Glucose Measurement from LOINC (required): Specifies the type of measurement using codes from the ValueSet for blood glucose measurements. |
![]() ![]() |
Σ | 1..1 | dateTime | Time of measurement |
![]() ![]() |
SΣC | 0..1 | Quantity | Measured blood glucose value |
![]() ![]() ![]() |
Σ | 1..1 | decimal | Numerical value (with implicit precision) |
![]() ![]() ![]() |
?!Σ | 0..1 | code | < | <= | >= | > - how to understand the value Binding: QuantityComparator (required): How the Quantity should be understood and represented. |
![]() ![]() ![]() |
ΣC | 1..1 | uri | System that defines coded unit form Fixed Value: http://unitsofmeasure.org |
![]() ![]() ![]() |
Σ | 1..1 | code | Coded form of the unit Binding: UCUMCodes (required): Defines the unit of measurement using codes from the UCUM units ValueSet. The UCUM code MUST be compliant with the unit that is linked with the LOINC code given as Observation.code. |
![]() ![]() |
1..1 | Reference(Device – Personal Health Device | DeviceMetric – Sensor Type and Calibration Status) | Reference to personal health device or its sensor type and calibration status | |
Documentation for this format | ||||
| Path | Status | Usage | ValueSet | Version | Source |
| Observation.status | Base | required | ObservationStatus | 📍4.0.1 | FHIR Std. |
| Observation.code | Base | required | ValueSet - Blood Glucose Measurement from LOINC | 📦0.1.0 | This IG |
| Observation.value[x].comparator | Base | required | QuantityComparator | 📍4.0.1 | FHIR Std. |
| Observation.value[x].code | Base | required | UCUM Codes | 📦4.0.1 | FHIR Std. |
| Id | Grade | Path(s) | Description | Expression |
| dom-2 | error | Observation | If the resource is contained in another resource, it SHALL NOT contain nested Resources |
contained.contained.empty()
|
| dom-3 | error | Observation | If the resource is contained in another resource, it SHALL be referred to from elsewhere in the resource or SHALL refer to the containing resource |
contained.where((('#'+id in (%resource.descendants().reference | %resource.descendants().as(canonical) | %resource.descendants().as(uri) | %resource.descendants().as(url))) or descendants().where(reference = '#').exists() or descendants().where(as(canonical) = '#').exists() or descendants().where(as(canonical) = '#').exists()).not()).trace('unmatched', id).empty()
|
| dom-4 | error | Observation | If a resource is contained in another resource, it SHALL NOT have a meta.versionId or a meta.lastUpdated |
contained.meta.versionId.empty() and contained.meta.lastUpdated.empty()
|
| dom-5 | error | Observation | If a resource is contained in another resource, it SHALL NOT have a security label |
contained.meta.security.empty()
|
| dom-6 | best practice | Observation | A resource should have narrative for robust management |
text.`div`.exists()
|
| ele-1 | error | **ALL** elements | All FHIR elements must have a @value or children |
hasValue() or (children().count() > id.count())
|
| ext-1 | error | **ALL** extensions | Must have either extensions or value[x], not both |
extension.exists() != value.exists()
|
| obs-6 | error | Observation | dataAbsentReason SHALL only be present if Observation.value[x] is not present |
dataAbsentReason.empty() or value.empty()
|
| obs-7 | error | Observation | If Observation.code is the same as an Observation.component.code then the value element associated with the code SHALL NOT be present |
value.empty() or component.code.where(coding.intersect(%resource.code.coding).exists()).empty()
|
This structure is derived from Observation
| Name | Flags | Card. | Type | Description & Constraints Filter: ![]() ![]() |
|---|---|---|---|---|
![]() |
0..* | Observation | Measurements and simple assertions | |
![]() ![]() |
1..1 | code | Measurement status Fixed Value: final | |
![]() ![]() |
1..1 | CodeableConcept | Measurement type of blood glucose Binding: ValueSet - Blood Glucose Measurement from LOINC (required): Specifies the type of measurement using codes from the ValueSet for blood glucose measurements. | |
![]() ![]() |
1..1 | dateTime | Time of measurement | |
![]() ![]() |
S | 0..1 | Quantity | Measured blood glucose value |
![]() ![]() ![]() |
1..1 | decimal | Numerical value (with implicit precision) | |
![]() ![]() ![]() |
1..1 | uri | System that defines coded unit form Fixed Value: http://unitsofmeasure.org | |
![]() ![]() ![]() |
1..1 | code | Coded form of the unit Binding: UCUMCodes (required): Defines the unit of measurement using codes from the UCUM units ValueSet. The UCUM code MUST be compliant with the unit that is linked with the LOINC code given as Observation.code. | |
![]() ![]() |
1..1 | Reference(Device – Personal Health Device | DeviceMetric – Sensor Type and Calibration Status) | Reference to personal health device or its sensor type and calibration status | |
Documentation for this format | ||||
| Path | Status | Usage | ValueSet | Version | Source |
| Observation.code | Base | required | ValueSet - Blood Glucose Measurement from LOINC | 📦0.1.0 | This IG |
| Observation.value[x].code | Base | required | UCUM Codes | 📦4.0.1 | FHIR Std. |
| Name | Flags | Card. | Type | Description & Constraints Filter: ![]() ![]() | ||||
|---|---|---|---|---|---|---|---|---|
![]() |
C | 0..* | Observation | Measurements and simple assertions Constraints: obs-6, obs-7 | ||||
![]() ![]() |
Σ | 0..1 | id | Logical id of this artifact | ||||
![]() ![]() |
Σ | 0..1 | Meta | Metadata about the resource | ||||
![]() ![]() |
?!Σ | 0..1 | uri | A set of rules under which this content was created | ||||
![]() ![]() |
0..1 | code | Language of the resource content Binding: CommonLanguages (preferred): A human language.
| |||||
![]() ![]() |
0..1 | Narrative | Text summary of the resource, for human interpretation This profile does not constrain the narrative in regard to content, language, or traceability to data elements | |||||
![]() ![]() |
0..* | Resource | Contained, inline Resources | |||||
![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() |
?! | 0..* | Extension | Extensions that cannot be ignored | ||||
![]() ![]() |
Σ | 0..* | Identifier | Business Identifier for observation | ||||
![]() ![]() |
Σ | 0..* | Reference(CarePlan | DeviceRequest | ImmunizationRecommendation | MedicationRequest | NutritionOrder | ServiceRequest) | Fulfills plan, proposal or order | ||||
![]() ![]() |
Σ | 0..* | Reference(MedicationAdministration | MedicationDispense | MedicationStatement | Procedure | Immunization | ImagingStudy) | Part of referenced event | ||||
![]() ![]() |
?!Σ | 1..1 | code | Measurement status Binding: ObservationStatus (required): Codes providing the status of an observation. Fixed Value: final | ||||
![]() ![]() |
0..* | CodeableConcept | Classification of type of observation Binding: ObservationCategoryCodes (preferred): Codes for high level observation categories. | |||||
![]() ![]() |
Σ | 1..1 | CodeableConcept | Measurement type of blood glucose Binding: ValueSet - Blood Glucose Measurement from LOINC (required): Specifies the type of measurement using codes from the ValueSet for blood glucose measurements. | ||||
![]() ![]() |
Σ | 0..1 | Reference(Patient | Group | Device | Location) | Who and/or what the observation is about | ||||
![]() ![]() |
Σ | 0..* | Reference(Resource) | What the observation is about, when it is not about the subject of record | ||||
![]() ![]() |
Σ | 0..1 | Reference(Encounter) | Healthcare event during which this observation is made | ||||
![]() ![]() |
Σ | 1..1 | dateTime | Time of measurement | ||||
![]() ![]() |
Σ | 0..1 | instant | Date/Time this version was made available | ||||
![]() ![]() |
Σ | 0..* | Reference(Practitioner | PractitionerRole | Organization | CareTeam | Patient | RelatedPerson) | Who is responsible for the observation | ||||
![]() ![]() |
SΣC | 0..1 | Quantity | Measured blood glucose value | ||||
![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations Slice: Unordered, Open by value:url | |||||
![]() ![]() ![]() |
Σ | 1..1 | decimal | Numerical value (with implicit precision) | ||||
![]() ![]() ![]() |
?!Σ | 0..1 | code | < | <= | >= | > - how to understand the value Binding: QuantityComparator (required): How the Quantity should be understood and represented. | ||||
![]() ![]() ![]() |
Σ | 0..1 | string | Unit representation | ||||
![]() ![]() ![]() |
ΣC | 1..1 | uri | System that defines coded unit form Fixed Value: http://unitsofmeasure.org | ||||
![]() ![]() ![]() |
Σ | 1..1 | code | Coded form of the unit Binding: UCUMCodes (required): Defines the unit of measurement using codes from the UCUM units ValueSet. The UCUM code MUST be compliant with the unit that is linked with the LOINC code given as Observation.code. | ||||
![]() ![]() |
C | 0..1 | CodeableConcept | Why the result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | ||||
![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc. Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() |
0..* | Annotation | Comments about the observation | |||||
![]() ![]() |
0..1 | CodeableConcept | Observed body part Binding: SNOMEDCTBodyStructures (example): Codes describing anatomical locations. May include laterality. | |||||
![]() ![]() |
0..1 | CodeableConcept | How it was done Binding: ObservationMethods (example): Methods for simple observations. | |||||
![]() ![]() |
0..1 | Reference(Specimen) | Specimen used for this observation | |||||
![]() ![]() |
1..1 | Reference(Device – Personal Health Device | DeviceMetric – Sensor Type and Calibration Status) | Reference to personal health device or its sensor type and calibration status | |||||
![]() ![]() |
C | 0..* | BackboneElement | Provides guide for interpretation Constraints: obs-3 | ||||
![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() |
C | 0..1 | SimpleQuantity(4.0.1) | Low Range, if relevant | ||||
![]() ![]() ![]() |
C | 0..1 | SimpleQuantity(4.0.1) | High Range, if relevant | ||||
![]() ![]() ![]() |
0..1 | CodeableConcept | Reference range qualifier Binding: ObservationReferenceRangeMeaningCodes (preferred): Code for the meaning of a reference range. | |||||
![]() ![]() ![]() |
0..* | CodeableConcept | Reference range population Binding: ObservationReferenceRangeAppliesToCodes (example): Codes identifying the population the reference range applies to. | |||||
![]() ![]() ![]() |
0..1 | Range | Applicable age range, if relevant | |||||
![]() ![]() ![]() |
0..1 | string | Text based reference range in an observation | |||||
![]() ![]() |
Σ | 0..* | Reference(Observation | QuestionnaireResponse | MolecularSequence) | Related resource that belongs to the Observation group | ||||
![]() ![]() |
Σ | 0..* | Reference(DocumentReference | ImagingStudy | Media | QuestionnaireResponse | Observation | MolecularSequence) | Related measurements the observation is made from | ||||
![]() ![]() |
Σ | 0..* | BackboneElement | Component results | ||||
![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() |
Σ | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. | ||||
![]() ![]() ![]() |
Σ | 0..1 | Actual component result | |||||
![]() ![]() ![]() ![]() |
Quantity | |||||||
![]() ![]() ![]() ![]() |
CodeableConcept | |||||||
![]() ![]() ![]() ![]() |
string | |||||||
![]() ![]() ![]() ![]() |
boolean | |||||||
![]() ![]() ![]() ![]() |
integer | |||||||
![]() ![]() ![]() ![]() |
Range | |||||||
![]() ![]() ![]() ![]() |
Ratio | |||||||
![]() ![]() ![]() ![]() |
SampledData | |||||||
![]() ![]() ![]() ![]() |
time | |||||||
![]() ![]() ![]() ![]() |
dateTime | |||||||
![]() ![]() ![]() ![]() |
Period | |||||||
![]() ![]() ![]() |
C | 0..1 | CodeableConcept | Why the component result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | ||||
![]() ![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc. Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() ![]() |
0..* | See referenceRange (Observation) | Provides guide for interpretation of component result | |||||
Documentation for this format | ||||||||
| Path | Status | Usage | ValueSet | Version | Source |
| Observation.language | Base | preferred | Common Languages | 📍4.0.1 | FHIR Std. |
| Observation.status | Base | required | ObservationStatus | 📍4.0.1 | FHIR Std. |
| Observation.category | Base | preferred | Observation Category Codes | 📍4.0.1 | FHIR Std. |
| Observation.code | Base | required | ValueSet - Blood Glucose Measurement from LOINC | 📦0.1.0 | This IG |
| Observation.value[x].comparator | Base | required | QuantityComparator | 📍4.0.1 | FHIR Std. |
| Observation.value[x].code | Base | required | UCUM Codes | 📦4.0.1 | FHIR Std. |
| Observation.dataAbsentReason | Base | extensible | DataAbsentReason | 📍4.0.1 | FHIR Std. |
| Observation.interpretation | Base | extensible | Observation Interpretation Codes | 📍4.0.1 | FHIR Std. |
| Observation.bodySite | Base | example | SNOMED CT Body Structures | 📍4.0.1 | FHIR Std. |
| Observation.method | Base | example | Observation Methods | 📍4.0.1 | FHIR Std. |
| Observation.referenceRange.type | Base | preferred | Observation Reference Range Meaning Codes | 📍4.0.1 | FHIR Std. |
| Observation.referenceRange.appliesTo | Base | example | Observation Reference Range Applies To Codes | 📍4.0.1 | FHIR Std. |
| Observation.component.code | Base | example | LOINC Codes | 📍4.0.1 | FHIR Std. |
| Observation.component.dataAbsentReason | Base | extensible | DataAbsentReason | 📍4.0.1 | FHIR Std. |
| Observation.component.interpretation | Base | extensible | Observation Interpretation Codes | 📍4.0.1 | FHIR Std. |
| Id | Grade | Path(s) | Description | Expression |
| dom-2 | error | Observation | If the resource is contained in another resource, it SHALL NOT contain nested Resources |
contained.contained.empty()
|
| dom-3 | error | Observation | If the resource is contained in another resource, it SHALL be referred to from elsewhere in the resource or SHALL refer to the containing resource |
contained.where((('#'+id in (%resource.descendants().reference | %resource.descendants().as(canonical) | %resource.descendants().as(uri) | %resource.descendants().as(url))) or descendants().where(reference = '#').exists() or descendants().where(as(canonical) = '#').exists() or descendants().where(as(canonical) = '#').exists()).not()).trace('unmatched', id).empty()
|
| dom-4 | error | Observation | If a resource is contained in another resource, it SHALL NOT have a meta.versionId or a meta.lastUpdated |
contained.meta.versionId.empty() and contained.meta.lastUpdated.empty()
|
| dom-5 | error | Observation | If a resource is contained in another resource, it SHALL NOT have a security label |
contained.meta.security.empty()
|
| dom-6 | best practice | Observation | A resource should have narrative for robust management |
text.`div`.exists()
|
| ele-1 | error | **ALL** elements | All FHIR elements must have a @value or children |
hasValue() or (children().count() > id.count())
|
| ext-1 | error | **ALL** extensions | Must have either extensions or value[x], not both |
extension.exists() != value.exists()
|
| obs-3 | error | Observation.referenceRange | Must have at least a low or a high or text |
low.exists() or high.exists() or text.exists()
|
| obs-6 | error | Observation | dataAbsentReason SHALL only be present if Observation.value[x] is not present |
dataAbsentReason.empty() or value.empty()
|
| obs-7 | error | Observation | If Observation.code is the same as an Observation.component.code then the value element associated with the code SHALL NOT be present |
value.empty() or component.code.where(coding.intersect(%resource.code.coding).exists()).empty()
|
This structure is derived from Observation
Key Elements View
| Name | Flags | Card. | Type | Description & Constraints Filter: ![]() ![]() |
|---|---|---|---|---|
![]() |
C | 0..* | Observation | Measurements and simple assertions Constraints: obs-6, obs-7 |
![]() ![]() |
?!Σ | 0..1 | uri | A set of rules under which this content was created |
![]() ![]() |
?! | 0..* | Extension | Extensions that cannot be ignored |
![]() ![]() |
?!Σ | 1..1 | code | Measurement status Binding: ObservationStatus (required): Codes providing the status of an observation. Fixed Value: final |
![]() ![]() |
Σ | 1..1 | CodeableConcept | Measurement type of blood glucose Binding: ValueSet - Blood Glucose Measurement from LOINC (required): Specifies the type of measurement using codes from the ValueSet for blood glucose measurements. |
![]() ![]() |
Σ | 1..1 | dateTime | Time of measurement |
![]() ![]() |
SΣC | 0..1 | Quantity | Measured blood glucose value |
![]() ![]() ![]() |
Σ | 1..1 | decimal | Numerical value (with implicit precision) |
![]() ![]() ![]() |
?!Σ | 0..1 | code | < | <= | >= | > - how to understand the value Binding: QuantityComparator (required): How the Quantity should be understood and represented. |
![]() ![]() ![]() |
ΣC | 1..1 | uri | System that defines coded unit form Fixed Value: http://unitsofmeasure.org |
![]() ![]() ![]() |
Σ | 1..1 | code | Coded form of the unit Binding: UCUMCodes (required): Defines the unit of measurement using codes from the UCUM units ValueSet. The UCUM code MUST be compliant with the unit that is linked with the LOINC code given as Observation.code. |
![]() ![]() |
1..1 | Reference(Device – Personal Health Device | DeviceMetric – Sensor Type and Calibration Status) | Reference to personal health device or its sensor type and calibration status | |
Documentation for this format | ||||
| Path | Status | Usage | ValueSet | Version | Source |
| Observation.status | Base | required | ObservationStatus | 📍4.0.1 | FHIR Std. |
| Observation.code | Base | required | ValueSet - Blood Glucose Measurement from LOINC | 📦0.1.0 | This IG |
| Observation.value[x].comparator | Base | required | QuantityComparator | 📍4.0.1 | FHIR Std. |
| Observation.value[x].code | Base | required | UCUM Codes | 📦4.0.1 | FHIR Std. |
| Id | Grade | Path(s) | Description | Expression |
| dom-2 | error | Observation | If the resource is contained in another resource, it SHALL NOT contain nested Resources |
contained.contained.empty()
|
| dom-3 | error | Observation | If the resource is contained in another resource, it SHALL be referred to from elsewhere in the resource or SHALL refer to the containing resource |
contained.where((('#'+id in (%resource.descendants().reference | %resource.descendants().as(canonical) | %resource.descendants().as(uri) | %resource.descendants().as(url))) or descendants().where(reference = '#').exists() or descendants().where(as(canonical) = '#').exists() or descendants().where(as(canonical) = '#').exists()).not()).trace('unmatched', id).empty()
|
| dom-4 | error | Observation | If a resource is contained in another resource, it SHALL NOT have a meta.versionId or a meta.lastUpdated |
contained.meta.versionId.empty() and contained.meta.lastUpdated.empty()
|
| dom-5 | error | Observation | If a resource is contained in another resource, it SHALL NOT have a security label |
contained.meta.security.empty()
|
| dom-6 | best practice | Observation | A resource should have narrative for robust management |
text.`div`.exists()
|
| ele-1 | error | **ALL** elements | All FHIR elements must have a @value or children |
hasValue() or (children().count() > id.count())
|
| ext-1 | error | **ALL** extensions | Must have either extensions or value[x], not both |
extension.exists() != value.exists()
|
| obs-6 | error | Observation | dataAbsentReason SHALL only be present if Observation.value[x] is not present |
dataAbsentReason.empty() or value.empty()
|
| obs-7 | error | Observation | If Observation.code is the same as an Observation.component.code then the value element associated with the code SHALL NOT be present |
value.empty() or component.code.where(coding.intersect(%resource.code.coding).exists()).empty()
|
Differential View
This structure is derived from Observation
| Name | Flags | Card. | Type | Description & Constraints Filter: ![]() ![]() |
|---|---|---|---|---|
![]() |
0..* | Observation | Measurements and simple assertions | |
![]() ![]() |
1..1 | code | Measurement status Fixed Value: final | |
![]() ![]() |
1..1 | CodeableConcept | Measurement type of blood glucose Binding: ValueSet - Blood Glucose Measurement from LOINC (required): Specifies the type of measurement using codes from the ValueSet for blood glucose measurements. | |
![]() ![]() |
1..1 | dateTime | Time of measurement | |
![]() ![]() |
S | 0..1 | Quantity | Measured blood glucose value |
![]() ![]() ![]() |
1..1 | decimal | Numerical value (with implicit precision) | |
![]() ![]() ![]() |
1..1 | uri | System that defines coded unit form Fixed Value: http://unitsofmeasure.org | |
![]() ![]() ![]() |
1..1 | code | Coded form of the unit Binding: UCUMCodes (required): Defines the unit of measurement using codes from the UCUM units ValueSet. The UCUM code MUST be compliant with the unit that is linked with the LOINC code given as Observation.code. | |
![]() ![]() |
1..1 | Reference(Device – Personal Health Device | DeviceMetric – Sensor Type and Calibration Status) | Reference to personal health device or its sensor type and calibration status | |
Documentation for this format | ||||
| Path | Status | Usage | ValueSet | Version | Source |
| Observation.code | Base | required | ValueSet - Blood Glucose Measurement from LOINC | 📦0.1.0 | This IG |
| Observation.value[x].code | Base | required | UCUM Codes | 📦4.0.1 | FHIR Std. |
Snapshot View
| Name | Flags | Card. | Type | Description & Constraints Filter: ![]() ![]() | ||||
|---|---|---|---|---|---|---|---|---|
![]() |
C | 0..* | Observation | Measurements and simple assertions Constraints: obs-6, obs-7 | ||||
![]() ![]() |
Σ | 0..1 | id | Logical id of this artifact | ||||
![]() ![]() |
Σ | 0..1 | Meta | Metadata about the resource | ||||
![]() ![]() |
?!Σ | 0..1 | uri | A set of rules under which this content was created | ||||
![]() ![]() |
0..1 | code | Language of the resource content Binding: CommonLanguages (preferred): A human language.
| |||||
![]() ![]() |
0..1 | Narrative | Text summary of the resource, for human interpretation This profile does not constrain the narrative in regard to content, language, or traceability to data elements | |||||
![]() ![]() |
0..* | Resource | Contained, inline Resources | |||||
![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() |
?! | 0..* | Extension | Extensions that cannot be ignored | ||||
![]() ![]() |
Σ | 0..* | Identifier | Business Identifier for observation | ||||
![]() ![]() |
Σ | 0..* | Reference(CarePlan | DeviceRequest | ImmunizationRecommendation | MedicationRequest | NutritionOrder | ServiceRequest) | Fulfills plan, proposal or order | ||||
![]() ![]() |
Σ | 0..* | Reference(MedicationAdministration | MedicationDispense | MedicationStatement | Procedure | Immunization | ImagingStudy) | Part of referenced event | ||||
![]() ![]() |
?!Σ | 1..1 | code | Measurement status Binding: ObservationStatus (required): Codes providing the status of an observation. Fixed Value: final | ||||
![]() ![]() |
0..* | CodeableConcept | Classification of type of observation Binding: ObservationCategoryCodes (preferred): Codes for high level observation categories. | |||||
![]() ![]() |
Σ | 1..1 | CodeableConcept | Measurement type of blood glucose Binding: ValueSet - Blood Glucose Measurement from LOINC (required): Specifies the type of measurement using codes from the ValueSet for blood glucose measurements. | ||||
![]() ![]() |
Σ | 0..1 | Reference(Patient | Group | Device | Location) | Who and/or what the observation is about | ||||
![]() ![]() |
Σ | 0..* | Reference(Resource) | What the observation is about, when it is not about the subject of record | ||||
![]() ![]() |
Σ | 0..1 | Reference(Encounter) | Healthcare event during which this observation is made | ||||
![]() ![]() |
Σ | 1..1 | dateTime | Time of measurement | ||||
![]() ![]() |
Σ | 0..1 | instant | Date/Time this version was made available | ||||
![]() ![]() |
Σ | 0..* | Reference(Practitioner | PractitionerRole | Organization | CareTeam | Patient | RelatedPerson) | Who is responsible for the observation | ||||
![]() ![]() |
SΣC | 0..1 | Quantity | Measured blood glucose value | ||||
![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations Slice: Unordered, Open by value:url | |||||
![]() ![]() ![]() |
Σ | 1..1 | decimal | Numerical value (with implicit precision) | ||||
![]() ![]() ![]() |
?!Σ | 0..1 | code | < | <= | >= | > - how to understand the value Binding: QuantityComparator (required): How the Quantity should be understood and represented. | ||||
![]() ![]() ![]() |
Σ | 0..1 | string | Unit representation | ||||
![]() ![]() ![]() |
ΣC | 1..1 | uri | System that defines coded unit form Fixed Value: http://unitsofmeasure.org | ||||
![]() ![]() ![]() |
Σ | 1..1 | code | Coded form of the unit Binding: UCUMCodes (required): Defines the unit of measurement using codes from the UCUM units ValueSet. The UCUM code MUST be compliant with the unit that is linked with the LOINC code given as Observation.code. | ||||
![]() ![]() |
C | 0..1 | CodeableConcept | Why the result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | ||||
![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc. Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() |
0..* | Annotation | Comments about the observation | |||||
![]() ![]() |
0..1 | CodeableConcept | Observed body part Binding: SNOMEDCTBodyStructures (example): Codes describing anatomical locations. May include laterality. | |||||
![]() ![]() |
0..1 | CodeableConcept | How it was done Binding: ObservationMethods (example): Methods for simple observations. | |||||
![]() ![]() |
0..1 | Reference(Specimen) | Specimen used for this observation | |||||
![]() ![]() |
1..1 | Reference(Device – Personal Health Device | DeviceMetric – Sensor Type and Calibration Status) | Reference to personal health device or its sensor type and calibration status | |||||
![]() ![]() |
C | 0..* | BackboneElement | Provides guide for interpretation Constraints: obs-3 | ||||
![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() |
C | 0..1 | SimpleQuantity(4.0.1) | Low Range, if relevant | ||||
![]() ![]() ![]() |
C | 0..1 | SimpleQuantity(4.0.1) | High Range, if relevant | ||||
![]() ![]() ![]() |
0..1 | CodeableConcept | Reference range qualifier Binding: ObservationReferenceRangeMeaningCodes (preferred): Code for the meaning of a reference range. | |||||
![]() ![]() ![]() |
0..* | CodeableConcept | Reference range population Binding: ObservationReferenceRangeAppliesToCodes (example): Codes identifying the population the reference range applies to. | |||||
![]() ![]() ![]() |
0..1 | Range | Applicable age range, if relevant | |||||
![]() ![]() ![]() |
0..1 | string | Text based reference range in an observation | |||||
![]() ![]() |
Σ | 0..* | Reference(Observation | QuestionnaireResponse | MolecularSequence) | Related resource that belongs to the Observation group | ||||
![]() ![]() |
Σ | 0..* | Reference(DocumentReference | ImagingStudy | Media | QuestionnaireResponse | Observation | MolecularSequence) | Related measurements the observation is made from | ||||
![]() ![]() |
Σ | 0..* | BackboneElement | Component results | ||||
![]() ![]() ![]() |
0..1 | string | Unique id for inter-element referencing | |||||
![]() ![]() ![]() |
0..* | Extension | Additional content defined by implementations | |||||
![]() ![]() ![]() |
?!Σ | 0..* | Extension | Extensions that cannot be ignored even if unrecognized | ||||
![]() ![]() ![]() |
Σ | 1..1 | CodeableConcept | Type of component observation (code / type) Binding: LOINCCodes (example): Codes identifying names of simple observations. | ||||
![]() ![]() ![]() |
Σ | 0..1 | Actual component result | |||||
![]() ![]() ![]() ![]() |
Quantity | |||||||
![]() ![]() ![]() ![]() |
CodeableConcept | |||||||
![]() ![]() ![]() ![]() |
string | |||||||
![]() ![]() ![]() ![]() |
boolean | |||||||
![]() ![]() ![]() ![]() |
integer | |||||||
![]() ![]() ![]() ![]() |
Range | |||||||
![]() ![]() ![]() ![]() |
Ratio | |||||||
![]() ![]() ![]() ![]() |
SampledData | |||||||
![]() ![]() ![]() ![]() |
time | |||||||
![]() ![]() ![]() ![]() |
dateTime | |||||||
![]() ![]() ![]() ![]() |
Period | |||||||
![]() ![]() ![]() |
C | 0..1 | CodeableConcept | Why the component result is missing Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing. | ||||
![]() ![]() ![]() |
0..* | CodeableConcept | High, low, normal, etc. Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations. | |||||
![]() ![]() ![]() |
0..* | See referenceRange (Observation) | Provides guide for interpretation of component result | |||||
Documentation for this format | ||||||||
| Path | Status | Usage | ValueSet | Version | Source |
| Observation.language | Base | preferred | Common Languages | 📍4.0.1 | FHIR Std. |
| Observation.status | Base | required | ObservationStatus | 📍4.0.1 | FHIR Std. |
| Observation.category | Base | preferred | Observation Category Codes | 📍4.0.1 | FHIR Std. |
| Observation.code | Base | required | ValueSet - Blood Glucose Measurement from LOINC | 📦0.1.0 | This IG |
| Observation.value[x].comparator | Base | required | QuantityComparator | 📍4.0.1 | FHIR Std. |
| Observation.value[x].code | Base | required | UCUM Codes | 📦4.0.1 | FHIR Std. |
| Observation.dataAbsentReason | Base | extensible | DataAbsentReason | 📍4.0.1 | FHIR Std. |
| Observation.interpretation | Base | extensible | Observation Interpretation Codes | 📍4.0.1 | FHIR Std. |
| Observation.bodySite | Base | example | SNOMED CT Body Structures | 📍4.0.1 | FHIR Std. |
| Observation.method | Base | example | Observation Methods | 📍4.0.1 | FHIR Std. |
| Observation.referenceRange.type | Base | preferred | Observation Reference Range Meaning Codes | 📍4.0.1 | FHIR Std. |
| Observation.referenceRange.appliesTo | Base | example | Observation Reference Range Applies To Codes | 📍4.0.1 | FHIR Std. |
| Observation.component.code | Base | example | LOINC Codes | 📍4.0.1 | FHIR Std. |
| Observation.component.dataAbsentReason | Base | extensible | DataAbsentReason | 📍4.0.1 | FHIR Std. |
| Observation.component.interpretation | Base | extensible | Observation Interpretation Codes | 📍4.0.1 | FHIR Std. |
| Id | Grade | Path(s) | Description | Expression |
| dom-2 | error | Observation | If the resource is contained in another resource, it SHALL NOT contain nested Resources |
contained.contained.empty()
|
| dom-3 | error | Observation | If the resource is contained in another resource, it SHALL be referred to from elsewhere in the resource or SHALL refer to the containing resource |
contained.where((('#'+id in (%resource.descendants().reference | %resource.descendants().as(canonical) | %resource.descendants().as(uri) | %resource.descendants().as(url))) or descendants().where(reference = '#').exists() or descendants().where(as(canonical) = '#').exists() or descendants().where(as(canonical) = '#').exists()).not()).trace('unmatched', id).empty()
|
| dom-4 | error | Observation | If a resource is contained in another resource, it SHALL NOT have a meta.versionId or a meta.lastUpdated |
contained.meta.versionId.empty() and contained.meta.lastUpdated.empty()
|
| dom-5 | error | Observation | If a resource is contained in another resource, it SHALL NOT have a security label |
contained.meta.security.empty()
|
| dom-6 | best practice | Observation | A resource should have narrative for robust management |
text.`div`.exists()
|
| ele-1 | error | **ALL** elements | All FHIR elements must have a @value or children |
hasValue() or (children().count() > id.count())
|
| ext-1 | error | **ALL** extensions | Must have either extensions or value[x], not both |
extension.exists() != value.exists()
|
| obs-3 | error | Observation.referenceRange | Must have at least a low or a high or text |
low.exists() or high.exists() or text.exists()
|
| obs-6 | error | Observation | dataAbsentReason SHALL only be present if Observation.value[x] is not present |
dataAbsentReason.empty() or value.empty()
|
| obs-7 | error | Observation | If Observation.code is the same as an Observation.component.code then the value element associated with the code SHALL NOT be present |
value.empty() or component.code.where(coding.intersect(%resource.code.coding).exists()).empty()
|
This structure is derived from Observation
Other representations of profile: CSV, Excel, Schematron